Infrared camera for scientific applications
The ThermaCAM S45 is a rugged (IP 54) infrared camera designed especially for Research and Development, scientific testing and general-purpose non-contact measurement. Hand-held or mounted on a tripod, operated locally or by remote control, with optional lenses and/or filters, the ThermaCAM S45 can be configured to meet a variety of scientific applications.
Precision Temperature Measurement And Extraordinary Image Quality
A state-of-the-art, uncooled microbolometer allows you to see temperature differences as small as 0.08°C in a range from -40°C to +2,000°C. It produces outstanding noisefree crisp high-resolution images (320 x 240 pixels), offering more than 76,000 individual measurement points per image at a refresh rate of 50/60 Hz.
IEEE-1394 Firewire Digital Output
Enhanced connectivity options include composite video and IEEE-1394 interface for super fast image and data transfer of real-time, fully radiometric 14-bit JPEG images.
Quick And Accurate Temperature Measurements
View thermal images in the high-resolution viewfinder. Optional optics, including wide angle and close-up lenses, make the ThermaCAM S45 a good choice for the most demanding scientific applications.
Easy-to-Operate
Conveniently located buttons and a joystick, control all the features of the camera. Functions like autofocus, freezing and storing images are just a button away.
LCD Data Panel
An LCD screen gives you information about the status of specific camera functions. It provides you with information on battery usage, storage capacity, communication status and more.
Flexible JPEG Image Storage
The ThermaCAM S45 saves images as fully radiometric JPEGs. These images are stored on a removable FLASH PC card (128 MB)
Real-Time Digital Storage and Analysis
The ThermaCAM S45 feeds real-time data directly to your PC for recording and analysis. Coupled with FLIR’s ThermaCAM Researcher™ software, the S45 allows in-depth, extensive thermal analysis including 50/60 Hz real-time digital recording and evaluation of high-speed events. Static images can be stored on the camera’s removable multimedia FLASH card.
|
| Imaging Performance |
Field of view / min focus distance
Spatial resolution (IFOV)
Thermal sensitivity
Image frequency
Focus
Electronic zoom function
Detector type
Spectral range
Digital image enhancement |
24°x18° /0.3 m (with 35 mm lens)
1.3 mrad
0.08°C at 30°C
50/60 Hz non-interlaced
Automatic or manual
2, 4 inerpolating
Focal Plane Array (FPA), uncooled microbolometer 320 x 240 pixels
7.5 to 13µm
Normal and enhanced |
|
|
Image Presentation
|
Video output
Viewfinder
External display
|
RS170 EIA/NTSC or CCIR/PAL composite video and
IEEE-1394 FireWire output
Built-in, high-resolution color LCD (TFT)
4” LCD with integrated remote control (optional) |
| |
|
| Measurement |
Temperature range
Accuracy
Measurement mode
Atmospheric transmission correction
Optics transmission correction
Emissivity correction
Reflected ambient temperature correction
External optics/window correction |
-40°C to +500°C, (-40°F to +2,732°F)
Up to +2,000°C (3632°F), optional
±2°C, ±2% of reading
Spot/manual (up to 10 movable). Area (circle or box, up to 10 movable), isotherm, line profile, Delta T
Automatic, based on inputs for distance, atmospheric temperatureand relative humidity
Automatic, based on signals from internal sensors
Variable from 0.1 to 1.0 or select from listings in pre-defined
materials list
Automatic, based on input of reflected temperature
Automatic, based on input of optics/window transmission and temperature |
| |
|
| Images Storage |
Type
File Formats
Voice annotation of images
|
Removable Flash-card (128 MB)
Standard JPEG, 14 bit measurement data included
30 sec. of digital voice “clip” stored together with the image |
| |
|
| Lenses (Optional) |
Field of view/min focus distance
Lens identification |
7°x5.3°/4 m (with 122 mm lens)
12°x 9°/1.2 m (with 71 mm lens)
45°x 34°/0.1 m (with 18 mm lens)
80°x 60°/0.1 m (with 9 mm lens)
200µm close-up (64 mm x 48 mm/150 mm)
100µm close-up (34 mm x 25 mm/80 mm)
50µm close-up (15 mm x 11 mm/19 mm)
18µm close-up (6 mm x 4 mm/7 mm)
Automatic
|
| |
|
| System Status Indicator |
| LCD Display |
Shows status of battery and storage media. Indication of power, communication and storage modes |
| |
|
| Battery System |
Type
Operating time
Charging system
External power operation
Power saving |
Li-Ion, rechargeable, field replaceable
2 hours continuous operation
in camera (AC adapter or 12 V from car) or 2 bay intelligent charger
AC adapter 110/220 V AC, 50/60 Hz or 12 V from car
(cable with Std plug: optional)
Automatic shutdown and sleep mode (user selectable) |
| |
|
| Environmental Specification |
Operating temperature range
Storage temperature range
Humidity
Encapsulation
Shock
VIbration |
-15°C to +50°C (5°F to 122°F)
-40°C to +70°C (-40°F to 158°F)
Operating and storage 10% to 95%, non-condensing
IP 54 IEC 529
Operational: 25G, IEC 68-2-29
Operational: 2G, IEC 68-2-6 |
| |
|
| Physical Characteristics |
|
Weight
Size
Tripod mounting
|
1.4 kg incl. Battery •
1.2 kg excluding battery
100mm x 120mm x 220mm (3.9”x4.7”x8.7”)
1/4” – 20 |
| |
|
| Interfaces |
|
FireWire
USB/RS-232
Remote control
|
IEEE-1394 FireWire output (standard)
Image, measurement and voice transfer to PC
Optional |
| |
|
ThermaCAM™ S45 includes:
- Ir Camera
- Carrying Case
- Lens Cap, Shoulder Strap, Hand Strap
- Manual
- Batteries (2)
- Battery Charger
- Power Supply Including Cable
- Video Cable RCA-Plug
- USB Cable
- Firewire Cable 4/4 and 4/6 pin
- Headset
- Thermacam Quickview™ Software
|
|
|
|